LDS3000

Compact Helium Leak Detector with ULTRATEST Technology™

Compact Design for Maximum Performance in Helium Leak Detection

The LDS3000 helium leak detector combines high accuracy and repeatability with short cycle times, making it the ideal choice for vacuum leak testing systems. Its compact design ensures seamless integration with minimal installation effort and space requirements. Enhance your operations with modern fieldbus systems for optimized performance.The LDS3000 is also available as LDS3000 AQ for the use in accumulation systems. The product family of the LDS3000 is the most versatile solution for automated testing sytems, among others in the automotive industry.INFICON also offers a wide variety of calibrated test leaks for almost every application. To simplify and speed up the order process, please use the  Order Form for customized calibrated leaks for testing system.

Helium Savings and Higher Efficiency with EcoBoost

The unique and patented EcoBoost feature enables the LDS3000 leak tester to detect smallest leaks, even on high and decreasing tracer gas background levels. Due to excellent instrument quality leading to extremely precise, repeatable and linear measurements, an intelligent algorithm predicts decreasing tracer gas backgrounds with high accuracy.With EcoBoost, the LDS3000 Helium Leak Detector can detect leaks early in the testing cycle and without the risk of missing small leaks. EcoBoost can also allow for the use of smaller, less energy consuming pumps and reduced helium consumption.

Features

  • Compact design allows individualized, custom integration into leak detection systems
  • Communicative diversity through a wide range of analog and digital interfaces
  • Considerably improved cabling
  • The new, patented routine allows calibration of the LDS3000 leak detector within 20 seconds.

Advantages

  • Compact design  for quick and easy integration
  • Reliable performance
  • Highest accuracy and repeatability in the market
  • Marke t leader in cycle time: Rapid detection of leak rates up to 1 x 10-11 mbar l/s in less than 1 second
  • PROFIBUS interface for simplified connectivity and interoperability, remote access and high data transfer efficiency. 
  • Digital interfaces, including RS232, RS485, USB 
  • Wide range of analog interfaces 
  • Minimal downtimes and reduced maintenance requirements
  • Fast calibration times, typically around 20 seconds
  • Optional touch screen display

Typical Applications

  • Evaporators, condensers, compressors
  • Battery cell housings
  • Battery pack trays and lids
  • Airbag gas generators and ignitors
  • Valves
  • Brake lines, fuel lines
  • Hydraulic components
  • Combustion Engines

Specifications

Min. detectable leak rate for Helium “ULTRA”-Modembar·l/s<1E-11
Min. detectable leak rate for Helium “FINE”-Modembar·l/s<5E-11
Min. detectable leak rate for Helium “GROSS”-Modembar·l/s<1E-9
Min. detectable leak rate for Helium “Sniffer”-Modembar·l/s<1E-7
Measuring units mbar·l/s, Pa•m³/s, Atm cm³, g/a, ppm
Max. inlet pressure “ULTRA”-Modembar0.2
Max. inlet pressure “FINE”-Modembar0.9
Max. inlet pressure “GROSS”-Modembar18
Response times<1
Digital inputs/outputs (if used with IO1000 Modul) 10 inputs / 8 outputs
Control inputVmax. 35 (PLC compatibel)
Ion Source 2 yttrium/iridium long-life cathodes (3 years guarantee)
lin/log recorder outputV0-10
Serial interface (with IO1000 / BM1000) RS232, RS485, Fieldbus
Dimensions (L x W x H)mm
in
330 x 240 x 280
13 x 9.45 x 11.1
Ambient temperature rangeC
F
10 – 45
50-113
Vacuum connections DN 16 KF, DN 25 KF
For the operation of the LDS3000 leak detector you need as basis one data cable and a IO1000 Module or BM1000-Modul